Liebherr WKt 4552 GrandCru User Manual

Page 26

Advertising
background image

26

The evaluation of air temperatures of the LK(U)v, LCv, LK(U)exv, LG(U)ex and LGPv models according to NF X 15-140 using a PT 100 temperature probe with
25 g copper thermal mass shows the following results:

Results according to NF X 15-140

LK(U)v, LCv, LK(U)exv, LG(U)ex and LGT laboratory appliances with Comfort-Controller

8,0 °C

7,5 °C

7,0 °C

6,5 °C

6,0 °C

5,5 °C

5,0 °C

4,5 °C

4,0 °C

3,5 °C

3,0 °C

2,5 °C

2,0 °C

t1

t2

t3

t4

t5

t6

t7

t8

t9

Probes

LKUv 1610

H0 = Gradient

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

8,0 °C

7,5 °C

7,0 °C

6,5 °C

6,0 °C

5,5 °C

5,0 °C

4,5 °C

4,0 °C

3,5 °C

3,0 °C

2,5 °C

2,0 °C

t1

t2

t3

t4

t5

t6

t7

t8

t9

Probes

LKUv 1612

H0 = Gradient

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

Gradient H0- :

4.0 K

(+2.0 K / – 2.0 K from

set temperature)

Max. fluctuation:

3.3 K

Gradient H0- :

2.9 K

(+1.3 K / –1.6 K from

set temperature)

Max. fluctuation:

2.5 K

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

Set temperature

Average probe temperature

Set temperature

Average probe temperature

Advertising