Liebherr WKt 4552 GrandCru User Manual

Page 32

Advertising
background image

32

LK(U)v, LCv, LK(U)exv, LG(U)ex and LGT laboratory appliances with Comfort-Controller

t1

t2

t3

t4

t5

t6

t7

t8

t9

Probes

LGT 4725 *

H0 = Gradient

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

Gradient H0-:

3,27 K

(+ 0,8 K/-2,47 K from

set temperature)

Max. fluctuation:

1,1 K

*If the cabinet is operated at -20 °C it is recommended to reduce the predefined offset of 2,7 K to 1,0 K (see user manual).

-42,0 °C

-42,5 °C

-43,0°C

-43,5°C

-44,0°C

-44,5 °C

-45,0 °C

-45,5 °C

-46,0 °C

-46,5 °C

-47,0 °C

-47,5°C

-48,0 °C

Set temperature

Average probe temperature

Advertising