Liebherr WKt 4552 GrandCru User Manual
Page 34
34
In order to comply with the DIN 58345 standard, the MKv and MKUv models have the following specifications:
1) To avoid any unwanted manipulation, the set temperature is fixed at +5 °C to guarantee maximum safety
2) Likewise the offset values of the probes cannot be changed
3) Also alarm limits cannot be changed from the default values +2/+8°C
4) An additional probe is positioned at the warmest point within the unit and this temperature is displayed on the controller
5) The MKv 3912 has a load line to prevent pharmaceuticals being placed at the very top of the compartment
6) In order to prevent product loss or detriment due to incorrect storage, an integrated evaporator cover keeps products
away from the cold surface of the evaporator
7) The integrated power failure alarm gives an acoustic warning immediately upon power failure
Otherwise the MKv and MKUv models have the same features as the LKv and LKUv equivalents.
Safety feature table
Pharmacy refrigerators
compliant to DIN 58345
MKv 3910 MKv 3912 MKUv 1610 MKUv 1612
Spark-free interior
No
No
No
No
High / Low temperature alarm
audio and visual
Yes
Yes
Yes
Yes
Door-open alarm, audio and visual
Yes
Yes
Yes
Yes
Power failure warning immediately
upon power failure l
Yes
Yes
Yes
Yes
Volt-free contact
for external alarm signal
Yes
Yes
Yes
Yes
RS 485 data transfer
interface
Yes
Yes
Yes
Yes
RS 485 / RS 232 converter with
data monitoring software
Accessory
Accessory
Accessory
Accessory
NTC product temperature probe
Accessory
Accessory
Accessory
Accessory
Internal alarm memory
Yes
Yes
Yes
Yes
Min. / Max.
temperature memory
Yes
Yes
Yes
Yes
+ 2 °C safety thermostat
Yes
Yes
Yes
Yes
Access port for e.g. PT 100 probe
Yes
Yes
Yes
Yes
Probe failure alarm
Yes
Yes
Yes
Yes
Pharmacy refrigerators compliant to DIN 58345
Pharmacy refrigerators compliant to
DIN 58345