Liebherr WKt 4552 GrandCru User Manual

Page 34

Advertising
background image

34

In order to comply with the DIN 58345 standard, the MKv and MKUv models have the following specifications:

1) To avoid any unwanted manipulation, the set temperature is fixed at +5 °C to guarantee maximum safety
2) Likewise the offset values of the probes cannot be changed
3) Also alarm limits cannot be changed from the default values +2/+8°C
4) An additional probe is positioned at the warmest point within the unit and this temperature is displayed on the controller
5) The MKv 3912 has a load line to prevent pharmaceuticals being placed at the very top of the compartment
6) In order to prevent product loss or detriment due to incorrect storage, an integrated evaporator cover keeps products

away from the cold surface of the evaporator

7) The integrated power failure alarm gives an acoustic warning immediately upon power failure

Otherwise the MKv and MKUv models have the same features as the LKv and LKUv equivalents.

Safety feature table

Pharmacy refrigerators
compliant to DIN 58345

MKv 3910 MKv 3912 MKUv 1610 MKUv 1612

Spark-free interior

No

No

No

No

High / Low temperature alarm
audio and visual

Yes

Yes

Yes

Yes

Door-open alarm, audio and visual

Yes

Yes

Yes

Yes

Power failure warning immediately
upon power failure l

Yes

Yes

Yes

Yes

Volt-free contact
for external alarm signal

Yes

Yes

Yes

Yes

RS 485 data transfer
interface

Yes

Yes

Yes

Yes

RS 485 / RS 232 converter with
data monitoring software

Accessory

Accessory

Accessory

Accessory

NTC product temperature probe

Accessory

Accessory

Accessory

Accessory

Internal alarm memory

Yes

Yes

Yes

Yes

Min. / Max.
temperature memory

Yes

Yes

Yes

Yes

+ 2 °C safety thermostat

Yes

Yes

Yes

Yes

Access port for e.g. PT 100 probe

Yes

Yes

Yes

Yes

Probe failure alarm

Yes

Yes

Yes

Yes

Pharmacy refrigerators compliant to DIN 58345