KD Scientific Legato 130 Syringe Pump User’s Manual

Publication 5617-008 Rev D

Setting the Force Limits

The maximum linear force of the syringe pump is 11 lbs

or 5 kgs. This force is sufficient to damage delicate, low

volume syringes, as well as many standard syringes if

pumping viscous fluids. The % of maximum force applied

by the pusher block can be changed.

In the set up screen, press

Force

.

Set the force level.

Press

Other Value

to enter a custom force %

Accept

the changes made by pressing

and go back to the Set Force screen.

Cancel

the changes made by pressing

and go back to the Set Force screen.

For a Custom Force %

Enter the precentage for custom force %

Accept

the changes made by pressing

and go back to the Set Force screen.

Cancel

the changes made by pressing

and go back to the Set Force screen.

NOTE: A minimum value of approximately

30% is required to overcome friction forces

in the mechanism. You may have to set the

force level to a higher value depending on

the backpressure/friction in your system.

Press Here

Press here to enter

custom force %