KD Scientific Legato 130 Syringe Pump User’s Manual

Publication 5617-008 Rev D

NOTE: When using triggers, the first step

in your program should define the trigger

starting level (High or Low). You may then

use the trigger command to change the

level as desired for your Program.

I/O (OUT) Trigger

The I/O (OUT) Trigger command is used to send a signal

to an external device when the Step in the Program is

reached.

The following options are available when setting an I/O

(OUT) Trigger command:

Trigger 1: Specifies output via the Trigger 1

(Pin 4) port. When selected, you must also specify

whether the signal to be sent is High Signal or

Low Signal.

High Signal: Sets the output signal to logic high.

Low Signal: Sets the output signal to logic low.

I/O (OUT) Trigger Advanced Options menu

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

TO INSERT AN I/O (OUT) TRIGGER COMMAND:

1.

From the Program Main screen, choose Step

Definition to enter the Step Definition screen.

2. Highlight (select once) the Program Step that you

want the I/O (OUT) Trigger to follow.

3. Choose the

button to access the

Program Steps Menu screen.

4. Choose

Advanced Options

to enter the Select

Advanced Options screen.

5. Choose the

I/O (OUT) Trigger

button to enter the

Trigger Step Definition Screen.

6. Set your Trigger settings by using the following

procedure:

a.

To send a Trigger signal to an external device via

Pin 4 select whether the signal will be High or Low

by selecting the appropriate

button (selection will highlight in white).

7.

Choose the Accept button to save your changes

and return to the Step Definition screen.