KD Scientific Legato 130 Series User Manual

Page 57

Advertising
background image

57

Operating the KDS Legato 130

K

D

S

L

e

g

a

to

1

0

0

S

e

ri

e

s

U

s

e

r’s

M
a

n

u

a

l

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

KD Scientific Legato 130 Syringe Pump User’s Manual

Publication 5617-008 Rev D

Digital I/O commands

INPUT

Reads and displays the trigger input port status.

Command format:

input

Query response:

<lf>[##:]Low.<cr>

<lf>[##]<prompt>

or:

<lf>[##:]High.<cr>

<lf>[##]<prompt>

OUTPUT

Sets the level on one of the output ports.

Command format:

output {1} {high|low}

SYNC

Sets the level on the sync port.

Command format:

sync {high|low}

Internal commands

STATUS

Displays the raw status for use with a controlling computer.

Command format:

status

The output is in three integer fields and one flag field, all

separated by spaces and terminated by a carriage return

/ linefeed pair. The first integer is the current rate in

femtoliters per second. The second integer is the infuse

time in milliseconds. The time has a granularity of 1

millisecond. The third integer is the infused volume in

femtoliters. All three values are for the current direction.

The flag field consists of six flags.

Flag one is the motor direction and will be “i or I”

if the pump is infusing and “w or W” if the pump is

withdrawing. If the letter is lower case, the pump

motor is idle. If upper case, the pump motor is

running.

Flag two is the limit switch status. If the infuse limit

switch was hit, “i or I” is displayed. If the withdraw

limit switch was hit, “w or W” is displayed. If no

limit switch was hit or the pump does not have limit

switches, “.” is displayed.

Flag three is the stall status and will be “S” if the

pump has stalled. Otherwise it will be “.”.

Flag four is the trigger input state and will be “T”

if high and “.” if low.

Flag five is the direction port state and will be “I”

for infuse and “W” for withdraw.

Flag six is the target reached status. It will be “T”

if the target time or volume was reached and “.”

if not.