KD Scientific Legato 130 Syringe Pump User’s Manual

Publication 5617-008 Rev D

Resume Run on Power Loss

If Resume Run on Power Loss is selected the pump will

power up running when power resumes.

In the Set-up screen,

select

Resume Run After Power Loss

Accept

the changes made and go back

to the previous screen before Set-up mode

was selected.

Cancel

and go back to the previous

screen before Set-up mode was selected.

NOTE: This feature is only valid in Infuse

Only or Withdraw Only modes. This feature

will only execute if the pump is running at

the time when power is lost, and was

running with no target set.

Press Here

Press Here

Setting and Viewing Pump Information

The information on the pump’s parameters are displayed

in the Pump Information Screen.

In the Set-up screen, select

Pump Information

The current information on the pump will be displayed.

• The Model type

• The Software Version

• The Maintenance Due

• The Current Pump Address

Press

Maintenance Reset

to reset the # of hours

for maintenance. This will reset it to 100 hours.

Confirm the Maintenance Reset by pressing the

Maintenance Reset

again.

Accept

the changes made by pressing

and go back to the Set-up screen.

Cancel

the changes made by pressing

and go back to the Set-up screen.

Press here to

reset the

maintenance

counter

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

NOTE: Elapsed Timers and Total Infused

Volume will be reset when power resumes.

Counters and timers information prior to

loss of power are lost