STEP-BY-STEP OPERATION EXAMPLE: USING A

CUSTOM PROGRAM

The following steps describe operation using a Custom

Program.

In the following description, we are assuming the use of a

Program that is already saved on the KDS Legato 130.

Note that if you make any changes to the Program’s

operating parameters and choose Accept on the page

where the changes are made, the changes automatically

become part of the stored Program for future use.

1.

From the Program Configuration screen,

the Program you wish to use should be displayed

in the Program Select button. If the Program is not

displayed, select the correct Program using the

procedure outlined in Running The Pump.

2. Choose the Syringe Select button to enter the

Syringe Selection screen.

3. Follow the instructions in the Syringe Selection

section of this Guide to complete the selection of a

syringe from the built-in syringe definition list, or to

create a new syringe definition. When finished,

choose the Accept button to save your changes, or

choose the Cancel button to cancel your changes,

and return to the Program Configuration screen.

4. Review the Pump Control Profile Steps described in

the Step Definition button. Note that the software

will display the first three Steps in the Program in

the main button area. If the Program uses more

than three Steps, the last Step in the Program will

be displayed in the box that is still part of the

button beneath the Step Definition button.

5. To review the Program Steps in detail or to

update/modify any of the stored Pump Control

Profile Parameters, choose the Step Definition button

to enter the Program Steps Summary Screen.

The name of the Program is listed at the right of the

screen, and the complete list of Steps is listed in a

series of buttons is on the left of the screen.

6. To review details for any individual Step in the

Program, select the Step you wish to review, then

press the Step button again to enter the Step

Definition screen for that Step.

7.

To edit the parameters for any Pump Control

Profile, follow the instructions supplied in the

applicable Pump Operation Profile description

contained in the Setting Program Parameters

section later in this Guide.

8. To return to the Program Steps Summary screen

when finished reviewing or editing the Step

parameter details, choose the Accept button to

save your changes or choose the Cancel button to

cancel any changes you made.

9. When you are finished reviewing the operating

details for all Steps, choose the Enter button to

return to the Program Configuration screen.

10. To preview the Program, choose the Preview

Program button, or to run the Program, choose the

Program Run button.

Step Definition Screen showing the list of Steps

currently defined for the Program