Operating the KDS Legato 130

K

D

S

L

e

g

a

to

1

3

0

U

s

e

r’s

M
a

n

u

a

l

KD Scientific Legato 130 Syringe Pump User’s Manual

Publication 5617-008 Rev D

I/O EVENT TRIGGERS

The KDS Legato 130 pumps permit easy connection to

external I/O devices via a variety of interfaces, including

a built-in 15-pin D-sub connector. The KDS Legato 130

software incorporates simple commands for

incorporating I/O controls into your pump Programs.

Integrated commands for control of general trigger

signals are integrated in the software. The following

procedure describes the use of the Advanced Options

I/O commands. Following the procedure is a description

of the selection menu options for each of the I/O

command types. Details for the electrical connection

controls and pin-out description for the 15-pin User I/O

Connector are provided in Appendix C of this manual.

The KDS Legato 130 manages trigger events through

specific pins on the I/O Connector port on the back of

the pump. The following details define

pin logic:

Trigger is communicated via Pin 4.

Event signals are communicated via Pin 2.