KD Scientific Legato 130 Syringe Pump User’s Manual

Publication 5617-008 Rev D

TROUBLESHOOTING

1. Infusion Accuracy

To ensure infusion accuracy always use new syringe(s)

and measure syringe bore diameter and enter actual

dimensions in millimeters (mm) using the Custom Syringe

entry option. Additionally, make sure that the guide rods

and lead screw are properly lubricated.

NOTE: It is recommended that you save

your programs periodically by uploading

them to a PC.

KDS Legato 130 Display Error Messages

2. Out of Range

A value was entered or encountered in a pump program

that was beyond the pump’s limits.

3. Power Failure Notification

If power to the pump is interrupted during operation, an

alarm will sound to alert the user to the interruption.

When power is restored, an onscreen message is

displayed to indicate the interruption.

4. Unit Won’t Power Up

Verify the the Power Source Switch on the rear of the

pump is set to DC-IN if powering through power supply

or RS-485 if powering in a chain.

The KDS Legato 130 Pump will notify you of a power failure with a message

when power is restored