Upgrading Legato 130 Series Software

1. Upload the latest software version to your desktop (format is filename.srec).

2. Disconnect all I/O devices and then connect the pump to PC using a USB cable.

3. On the Set up screen of the pump, press the Upgrade Software button twice to enter “Boot Loader

Mode”.

4. From the CD provided with the pump, open the Legato 130 Series Pump Updater Application.

5. Click Open Update File and browse to the file previously saved on your desktop and click Open.

6. Click Start Update.

7. After the update is complete, press Reboot to power cycle the pump.