KD Scientific Legato 130 Syringe Pump User’s Manual

Publication 5617-008 Rev D

Exporting A Program To An

External Computer

Follow these instructions to export a Program to an

external computer.

1.

Connect the pump to the external computer via

a USB cable.

2. On the computer, open the HyperTerminal

application and verify that the computer can

communicate with the pump.

3. At the pump, from the Program Main screen,

choose Program Select then choose the Program

you would like to export.

4. Choose the

button.

5. In HyperTerminal, choose Transfer

→Capture Text...

6. Enter the File Name that you want to save

the Program under, then choose Start.

7.

Choose Export Program to PC.

8. The pump will display a confirmation message in

the Message Area on the screen when the export

is complete. Note that long Programs may take a

few seconds to export.

9. At the destination computer, in HyperTerminal,

when the Program is done transferring, the

message “mend” will appear onscreen.

10. In HyperTerminal, choose Transfer

→Capture

Text...

→Stop. Your Program is now saved on the

external computer.

‘FILE OPTIONS’

IMPORTING A PROGRAM FROM AN EXTERNAL

COMPUTER

Follow these instructions to import a Program from an

external computer.

1.

Connect the pump to the external computer via a

USB cable and set the Address value to

‘00’ using the Set Pump Address command in the

Settings menu.

2. On the computer, open the HyperTerminal

application and verify that the computer can

communicate with the pump.

3. In HyperTerminal, click on the Properties icons and

choose the Settings tab on the pop-up box that is

displayed.

4. Click on the ASCII Setup button.

5. In the Line Delay field, enter 500. This will set a

500 millisecond delay after each line is transferred

during the import process.

6. Click OK twice to return to the HyperTerminal

Main Menu screen.

7.

In HyperTerminal, choose Transfer

→Send Text

File...

8. Select the Program that you want to import to the

pump and click Open.

9. Before disconnecting, verify that the Program was

transferred by choosing Program Select from the

Program Main screen at the destination pump,

then browse the Program menu to confirm that the

new Program is now saved at the pump.

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

NOTE: The KD Scientific Syringe Pump Communicator program, provided on the Legato CD,

can be used to simplify the import / export process for programmable pumps. Please refer to

the Syringe Pump Communicator manual for instructions.