2 interactance immersion configuration, Interactance immersion configuration – Metrohm NIRS XDS Interactance OptiProbe Analyzer User Manual

Page 30

Advertising
background image

28

▪▪▪▪▪▪▪

Reference Standardization requires use of an XC-
1200 Calibrated 99% Reflectance Reference.

Do not attempt to perform this with the R99 from
the XC-1010 Standard Set. Please see the section
on Reference Standardization for full information.

When instructed to scan the Wavelength Standard,
or the Calibrated Reference Standard, simply press
the door, and the holder will open.

Pull as far as it will come out, then insert the cup,
window side toward the probe.

Close door and insert the probe into the middle
aperture, against the sample cup.

5.2

Interactance Immersion Configuration

This is the Interactance Immersion (Liquid Tip) used
to sampling liquids such as alcohols and other
clear to semi-clear liquids. The path length of the
tip is set by use of a spacer, as shown in a later
frame.

Select the proper Tip type in the Data Collection
Method (DCM) in Vision.

(See section 6.0 for full information.)