Metrohm NIRS XDS Interactance OptiProbe Analyzer User Manual

Page 62

Advertising
background image

60

▪▪▪▪▪▪▪

If the test has passed, the user should click the cursor on the tabular display, go to the File menu, and
click Print to keep a copy of the report. It is saved in the Diagnostic Database for future recall. See the
Vision manual for a full explanation.

Wavelength Certification also tests certain peaks in the wavelength standard. These peaks are used to
set the “Instrument Wavelength Profile,” and one peak is used for bandwidth calculation.

The wavelengths used for the instrument wavelength profile are well-defined, stable peaks in the
wavelength standard. These are the same peaks used during Instrument Calibration. Wavelength
Certification is a verification that the peaks are in correct positions, and that the peak positions are