Metrohm NIRS XDS Interactance OptiProbe Analyzer User Manual

Page 35

Advertising
background image

▪▪▪▪▪▪▪

33

6.

Using a large flat-bladed screwdriver,

loosen the large (3/4”) threaded plug inside
the instrument opening.

A larger screwdriver than the one shown is
preferable. A small screwdriver is shown to
give the best view of the threaded plug.

When the plug is loose, hold the
screwdriver blade horizontal (as shown) to
lift the threaded plug out of the unit.

7.

Remove the plug.

8.

The white object inside the opening is the

reflectance standard.

9.

Reach in with a tissue, then press the

reflectance standard out from the other
side -- insert the OptiProbe gently into the
reference position. The standard should
pop out into the tissue.

10.

Gently wipe the surface with a laboratory

tissue to clean it. Do not abrade or pit the
surface in any way.