7.2.1

Performance Test

Performance Test is a comprehensive test of instrumental performance, and is the final assurance that
the instrument is ready to run samples. The key items verified during this test are:

Instrument Noise in each of four wavelength regions

Internal Wavelength Performance (wavelength positions on non-traceable, internal reference

materials.)

Internal Wavelength Precision (Repeatability)

NIR Gain

Visible Gain

The test is initiated as follows:

Select Performance Test from the
Diagnostics menu. Click on Run
Performance Test.

Verify that the OptiProbe is inserted into the Reference
position as shown.

As the test runs, a screen like that shown on the next page is displayed. Upon test completion, a
message box is displayed to indicate test completion and status.