1 performance test, Performance test – Metrohm NIRS XDS Interactance OptiProbe Analyzer User Manual
Page 57
▪▪▪▪▪▪▪
55
7.2.1
Performance Test
Performance Test is a comprehensive test of instrumental performance, and is the final assurance that
the instrument is ready to run samples. The key items verified during this test are:
•
Instrument Noise in each of four wavelength regions
•
Internal Wavelength Performance (wavelength positions on non-traceable, internal reference
materials.)
•
Internal Wavelength Precision (Repeatability)
•
NIR Gain
•
Visible Gain
The test is initiated as follows:
Select Performance Test from the
Diagnostics menu. Click on Run
Performance Test.
Verify that the OptiProbe is inserted into the Reference
position as shown.
As the test runs, a screen like that shown on the next page is displayed. Upon test completion, a
message box is displayed to indicate test completion and status.