Metrohm NIRS XDS Interactance OptiProbe Analyzer User Manual

Page 46

Advertising
background image

44

▪▪▪▪▪▪▪

Vision acquires spectra needed to perform Wavelength
Linearization.

No operator action is required at this point – this screen is
for information only.

Vision continues.

No operator action is required at this point – this screen is
for information only.

Vision continues.

No operator action is required at this point – this screen is
for information only.

Once Vision has finished with this screen, a result screen will
be displayed.

A typical result
screen is shown.
Peak positions for
the reference
materials are
located using a
peak-finding
algorithm. These
“found” peaks are
compared to the
nominals.
Differences should
be no more than
0.4nm for any peak.

Click “Yes” to send the linearization to the instrument.

After the linearization is successfully sent to the instrument,
this message confirms the transfer.

Click “OK” to proceed.

This procedure is done twice, once for each direction of the grating motion. Repeat steps as
prompted by Vision.