Instrument cooling fans not
operating.

1.

Instrument may not be warmed up. Fans do not

operate until the instrument is near operating
temperature.

2.

Ambient temperature may not require fan cooling.

This is common in cool environments.

To check fan operation, cycle the Power Switch.
Note if the fans come on for a short burst,
approximately 25 seconds after power-up. If fans
come on, they are operating properly.

10

Software “freezes” or
“lockups” during use

Usually caused by use of two network cards in PC.
Data collisions are causing lost communication
“packets”. Remove one NIC card, and use the
connection method on page 4.

11

Vision reports that there is no
Reference Standardization or
Instrument Calibration;
suggests that these tests be
run.

1.

Usually caused by entering the test, then

canceling, which leaves the instrument with no
completed test constants. Do not cancel when
these tests have been initiated.

2.

May be caused by use of two network cards in PC.

Data collisions are causing lost communication
“packets”. Remove one NIC card, and use the
connection method on page 4.

If these measures do not correct the problem, please contact: