1 wavelength linearization, Wavelength linearization – Metrohm NIRS XDS Interactance OptiProbe Analyzer User Manual

Page 44

Advertising
background image

42

▪▪▪▪▪▪▪

This is used only on certain XDS Process Instruments. Do not select for XDS OptiProbe.

When method transfer (between instruments) is anticipated, be sure that Reference
Standardization and Instrument Calibration are selected. These features assure best method
transfer between similar instruments.

Some of these selections are spectroscopic
calibrations, and are used to apply corrections to the
instrument to minimize differences between units of
the same configuration.

The correction programs are accessed from the
Diagnostics menu bar, shown at right. The
diagnostic steps are explained in the sections that
follow.

7.1.1

Wavelength Linearization

Wavelength Linearization uses an internal wavelength standard set to determine a set of internal,
arbitrary peak positions that the instrument will use to maintain repeatability of wavelength response.

The NIR wavelength positions of
these peaks appear as shown.

The scale of this display is marked
in encoder pulses, which do not
relate to nanometers directly.

From the peaks, a linearization is
performed, which allows
assignment of nanometer values.