Advanced Entry Calibration (continued)

Step

Operator Action

Displayed Message

Comment

6

Press the

key

to scroll to ‘Advanced
Entry’.



You are entering

Advanced Entry
Calibration.

7

Press the

key.



Press the
key to abort this

procedure.

8

Press the

key.



You are beginning
the Advanced
Entry Calibration.

9

Press the

key.

Place the label under the
Sensor Eye Mark and
lower the printhead
assembly.


-OR-

If using preprinted
media, ensure the
label area under
the sensor is free
of preprinted text,
graphics or
borders.

10

Press the key
to select the setting
(denoted by an ‘*’) and
then record the sensor
reading (the ‘yyy’ value)
in a table similar to the
one shown after Step 12,
as a Label Value.

Next, press the
key to increment the
Gain Number.

Repeat this step for each
of the Gain Numbers
(00-32).


-OR-

Where ‘yyy’ is a
numerical value
representing the
current sensor
reading: useful in
locating the best
sensor position.

þ

Note: Never

position the
sensor over a
perforation when
recording a
sensor reading.