Advanced Entry Calibration (continued)

Step

Operator Action

Displayed Message

Comment

16

Press the

key.


Using the or the

key, set the

‘Paper’ level to the value
determined in the previous
step. Press the

key

to set the entry (indicated
by the ‘*’) and advance the
menu.

Repeat for the ‘Gap’ (or
‘Mark’) and ‘Empty’ levels.


в

в

Or, for reflective media:

в

в

The selection
will flash and
the display
heading will
change to
indicate the
item for entry.

17

After all entries have been
made, press the
key to back out of the menu
and then press the

key

to save the settings and

return to the Ready Mode.

From ‘Ready’,
press the FEED
key to advance
to the next label
TOF.


þ

Note: If the Advanced Entry Calibration fails, try the following procedure.

Re-enter Media Settings / Calibration / Advanced Entry / Sensor Gain
and lower the selected Gain Setting number by one (to make the
sensor less sensitive). Select the new Gain Setting, save the changes,
exit the menu, and then test your media. Repeat the procedure until a
usable media setting is obtained.