Brady 3481 User Manual

Page 78

Advertising
background image

68

Advanced Entry Calibration (continued)

Step

Operator Action

Displayed Message

Comment

12

From the data collected in
Steps 10 and 11, where
both the sensor readings are
above 20, subtract each
Label Value from the
corresponding TOF Value.
These are the Difference
Values
.

From the resulting list, find
the largest Difference
Value (see example below).
Its associated Gain Number
will provide the best
algorithm for your media.


Or, for reflective media:

Both sensor
readings must
be above 20.

Gain

Number

Label
Value

TOF

Value

Difference

Value

00

255

254

1

01

251

240

11

02

241

213

28

03

231

182

49

04

219

150

69

05

212

119

93

06

200

88

112

07

189

58

131

08

178

32

146

09

167

19

N/A

10

156

17

N/A

11

146

16

N/A

12

136

15

N/A

31

116

14

N/A

For example, if your
compiled data had the
values shown in this
table, Gain Number 8
would be chosen
because it has the
highest Difference
Value (146) where
both the Label Value
and the TOF Value
are above 20.

32

112

14

N/A