ADS Enigma/Enigma-Compact/Enigma-hyQ User Manual

13

Note: Hydrophone loggers cannot be inserted into the case with
hydrophones connected. Reconnect the hydrophone to the logger after
programming and before the first epoch.

The following table provides instructions and descriptions for the controls
involving programming and readback:

Control

Use/Indication

Description

Green
Button

Short Press

Reports status of last case programming
operation

Green
Button

Long Press

Starts programming sequence

Red Button

Short Press

Reports status of last case readback
operation

Red Button

Long Press

Starts readback sequence

LEDs (8)

On

Indicates success of appropriate
programming/ readback

LEDs (8)

Flash

Indicates logger is detected but failed to
program/readback

LEDs (8)

Off

Indicates no logger is detected

Note: The Enigma–Compact has 3 loggers and LEDs.

Programming

Program the loggers using the case in the following way:

1. Make sure the loggers to be programmed are in the case.

2. Press and hold the green button for about 1 second. Check for the

“walking LEDs” and fast beeping, and then close the case. The
beeping should slow down to about 1 beep per second after
approximately 8 seconds. This indicates that the case is programming
the loggers.

3. After hearing a long beep, open the case.

4. Verify whether the LED for each logger is on, indicating successful

programming. Please note that logger status is displayed for only 8
seconds. To re-display status information, follow the instructions in
the following section, Checking Status - Programming.

If the case fails to program a particular logger, clean the top of the logger and
the appropriate optical window in the top of the case and then repeat the
programming procedure.