3 transfer using the trans-blot, Turbo, System – Bio-Rad Mini-PROTEAN® TGX™ Precast Gels for 2-D Electrophoresis User Manual

Page 32

Advertising
background image

26

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

Foam pad

Membrane
Gel

Filter paper

Filter paper

Foam pad

Assembly of the Mini-Trans-Blot cassette .

11.2.3 Transfer Using the Trans-Blot

®

Turbo

System

1. Open the transfer pack and assemble the components on the cassette in the order shown. For best

results, use the roller to remove any air trapped between the layers. If using a single mini or midi
sandwich, place the sandwich in the middle of the cassette bottom. With two mini gels, place the
sandwiches on a midi stack with the foot of each gel facing the center of the stack.

2. Place the lid on the cassette and lock the lid in place by turning the knob clockwise, using the

symbols on the lid as a guide. Slide the cassette into the appropriate bay of the Trans-Blot Turbo cell.
Each cassette and bay can hold up to two mini gels or one midi gel (Table 11.1).

3. Start the transfer. With the cassette in the cell, press TURBO and select the gel type. Press A:RUN

or B:RUN to begin the transfer. Press LIST to select a Bio-Rad optimized protocol (Table 11.2) or a
user-defined protocol. Press NEW to create and run a new protocol.

4. When transfer completes, RUN COMPLETE appears. Pull the cassette straight out of the slot and

unlock the lid. Disassemble the blotting sandwich.

Refer to the Trans-Blot Turbo Instruction Manual (bulletin 10020688) for complete instructions.

Top ion reservoir stack

Gel

Cassette top

(–) electrode (cathode)

Membrane

Cassette bottom

(+) electrode (anode)

Bottom ion reservoir

stack

Assembly of the gel blot sandwich with the Trans-Blot Turbo system .

Mini-PROTEAN Precast Gels

Advertising