Instrument specifications, Reportable range, Accuracy – ITC Avoximeter 1000E Whole Blood Oximeter User Manual

Page 17: Precision, Interference

Advertising
background image

ITC AVOXimeter 1000E Operator’s Manual

11

Instrument Specifications

Size

20.3 cm (8.0 in) x 25.4 cm (10.0 in) x 9.5 cm (3.8 in)

Weight

1.8 kg (4 lbs)

Operating Temperature

Room temperature (15°C to 30°C, 59°F to 86°F)

Battery Type

Nickel Cadmium (NiCad)

Operating Time On Battery

Approximately 8 hours (constant run at medium brightness).
Tests may also be run while the ITC AVOXimeter 1000E is
plugged into the AC/DC power module.

Anticipated Battery Life

Approximately 500 charge / discharge cycles

Power Supply/Chargers

Input:

100 / 240 VAC, 50 / 60 Hz

Output: 12 VDC, 830 mA

Serial Data Port

RS232C

Sample Type

Whole blood

Sample Volume

50 µL

Analysis Time

7 to 10 seconds per sample

Analysis Wavelengths

5

Reportable Range

%HbO

2

0 to 100%

THb

4 to 25 g/dL

[O

2

Content]

0 to 35 mL O

2

/dL

Accuracy

%HbO

2

±1 %HbO

2

THb (>10 g/dL)

±0.45 g/dL

THb (<10 g/dL)

±0.35 g/dL

Precision

%HbO

2

0.5 %HbO

2

THb

0.3 g/dL

Interference

Bilirubin

None

Hemolysis

None

Carboxyhemoglobin

None

Methemoglobin

< 1% %HbO

2

, < 0.2 g/dL THb

(THb = 16 g/dL, MetHb < 10%, 7.1 < pH < 7.8)

Fetal Hemoglobin

< 1% %HbO

2

, < 0.45 g/dL THb

(THb = 16 g/dL, HbF < 100%)