22

Requiring Entry of a Patient ID

The instrument can be configured so that a Patient ID must be entered before a test can be run.

Important:

The User ID for the “QA User” must be entered in order to enable
the requirement to enter a Patient ID. See page 21 fo
r details.

1.

Display the “User and Patient ID” menu (a submenu of

, “Data Management,”

see page 8).

2.

Press

followed by

to display the current Patient ID setting:

3.

Press

followed by

. A screen is displayed for specifying the minimum

and maximum number of digits:

4.

If the minimum and/or maximum number of digits are to be changed, press

and enter a new value (followed by

) for the minimum and maximum

number of digits.

Note:

If the minimum and/or maximum number of digits is changed, the
instrument will continue to prompt that up to twelve digits is acceptable
when entering a Patient ID. Operators should be instructed to ignore
this prompt.

5.

Respond

to the confirmation prompt. The “User and Patient ID” menu is again

displayed.

6.

Press

to return to the previous menu, if desired.