ITC AVOXimeter 1000E Operator’s Manual

57

Symptom

Possible Cause

Action Required

THb and/or %HbO

2

readings for yellow or
orange filter are out of
range.

1. The optical detector may be

contaminated by blood or
foreign material.

1. Clean the optical detector (see

page 59).

2. The instrument may require

calibration.

2. Re-calibrate the instrument

(see page 39).

The following ERROR:
message is displayed:

“Uninitialized Vector
Service Required“

1. Microprocessor error.

1. Press the

and

keys

simultaneously to turn the
instrument off, then turn the
instrument back on.

2. If the problem persists, contact

ITC Technical Support (see
page ii).