ITC AVOXimeter 1000E Operator’s Manual

38

Enabling QC Lockout

If QC Lockout is enabled, the operator cannot run a test if optical quality control is not
successfully run at the specified intervals.

Note:

The User ID for the “QA User” must be entered in order to enable QC Lockout and
specify the interval for running optical quality control. See page 21 for details.

1.

Display the “Data Management” menu (a submenu of

, see page 8).

2.

Press

followed by

to display the current QC Lockout setting:

3.

Press

followed by

. A screen is displayed for specifying the interval

between optical quality control tests:

4.

To change the interval, press

and enter a new value (followed by

). The

“Data Management” menu is again displayed.

5.

Press

to return to the previous menu, if desired.