Replacing the battery – ITC Avoximeter 1000E Whole Blood Oximeter User Manual

Page 69

Advertising
background image

ITC AVOXimeter 1000E Operator’s Manual

63

Replacing the Battery

Materials Needed

:

1. # 0 Phillips screwdriver

2. Diagonal cutting pliers

3. Replacement battery with cable (E4-BATP)

4. Securing cording (packaged with replacement battery)

1. Steps

1 through 11 of the Cleaning the Optical Detector procedure on page

59 Must be performed before moving to step 2 of “Replacing the Battery”,
below.

Note:

The battery is housed in the lower cover.

2.

Using the diagonal cutting pliers, cut the plastic ties that secure the battery to the
frame.

Note:

Be sure to cut only the white plastic ties that are wrapped around the
battery and secure it to the lower cover. Do not cut the blue and white
power cables that are connected to the top of the battery.

3.

Remove the battery, being careful not to snag any components when pulling the
cable out from underneath the main circuit board. If the battery well contains
tape, remove the tape.

4.

Remove the liner from the tape on the bottom of the new battery, and place the
new battery in the battery well with the tape side down, ensuring that the blue and
white cables are facing up, then route the new plastic ties (packaged with the
replacement battery) underneath the battery well.

Plastic Tie

Plastic Tie

Blue and White
Battery Cable

New Plastic Tie

New Battery

New Plastic Tie

Advertising