ITC AVOXimeter 1000E Operator’s Manual

64

5.

Secure the new battery in the battery well, using the new white plastic ties.

Note:

When securing the battery to the battery well, be sure that the plastic ties
are routed underneath the battery well. Cut the excess length from the
plastic ties using the diagonal cutting pliers.

6.

Thread the blue and white battery cable of the new battery under the main circuit
board.

7.

Re-connect the cables, replace the upper cover of the ITC AVOXimeter 1000E,
connect the AC adapter, and turn on the instrument as outlined in Steps 17
through 25 of the Cleaning the Optical Detector procedure on page 59. The self-
test should run and the “Ready – Insert Cuvette” screen should be displayed.

Note:

Be sure to connect the blue and white battery cable last.

The new battery is not charged. To ensure adequate charge, leave the
instrument connected to the AC Adapter for a minimum of eight hours.
Tests can be run while the instrument is charging.

8.

Reset the time and date (see page 16).