Calibration, Cuvette pathlength, Re-calibration – ITC Avoximeter 1000E Whole Blood Oximeter User Manual

Page 45

Advertising
background image

ITC AVOXimeter 1000E Operator’s Manual

39

Calibration

The ITC AVOXimeter 1000E is factory-calibrated and employs highly stable state-of-the-art light
sources.

Cuvette Pathlength

Accuracy of total hemoglobin measurements depends on using the correct cuvette pathlength
(see page 1). The user must check the pathlength whenever using a different lot number of
cuvettes.

Note:

The cuvette lot number and pathlength are included on the carton label and bag for
each lot of cuvettes.

Entering a Different Cuvette Pathlength:

1.

Display the “Calibration” menu (a submenu of

, see page 8).

2.

Press

followed by

to display the screen for entering the cuvette

pathlength:

3.

To change the interval, press

and enter a new value:

Note:

Pathlengths ranging from 90 microns to 132 microns can be entered.

4.

To change the interval, press

and enter a new value.

5.

Check that the new pathlength is correctly displayed, then press

. The

“Calibration” menu is again displayed:

6.

Press

to return to the previous menu, if desired.

Re-Calibration

If the ITC AVOXimeter 1000E results are out of range with either blood samples or controls and
troubleshooting does not resolve the issue, re-calibration may be necessary. Please contact ITC
Technical Support for more information on re-calibration.