ITC AVOXimeter 1000E Operator’s Manual

37

Tagging a Liquid Control Test with a Lot Number

The lot number of the control being tested can be included in the results for the test.

Note:

Lot numbers of liquid controls must be entered before they can be tagged to a liquid
QC test (see page 36).

1.

Insert a cuvette filled with liquid quality control material (see page 35). A prompt to
confirm liquid QC is displayed:

2.

Press

to confirm. A prompt to confirm that the level and lot numbers are being

entered is displayed:

3.

Press

to confirm. A menu for selection of a control level is displayed:

4.

Select the level that is being run and press

. A menu for selection of a lot

number is displayed:

5.

Select the previously entered lot number that is being run and press

.

6.

The sample results will be displayed on the screen, and the lot number and LQC
level will be added to the record for that sample.