3-11

Setup and Programming

Clear All Memory Locations

Use the procedure in this section to clear all memory locations.

1. Press RECALL DATA.

— LCD displays as shown in Figure 3-5. The present

memory location at the upper left, the total number

of memory locations to the right and the time data

was stored at far right.

— LED display shows the measurement value of that

memory location while the LCD displays its

engineering units at the lower right.

2. Press MANUAL CLEAR.

— LCD display shows "CLEAR" and the present (left)

and total (right) memory locations on the upper line,

as shown in Figure 3-6.

3. Press right DIGIT twice to select "All" on the lower LCD

line.

4. Press ENTER to clear the entire data memory. LED

display reads:

NO DATA

The audible alarm sounds momentarily. The 2000-610-

02 automatically returns to measurements mode.

Setting Up Printer/PC Ports

The MULTITEST Indicator is equipped with two RS-232 serial

communications ports for use with an external printer or

personal computer. The transfer rate is programmable from 300

to 19.2K baud rate with 8 data bits, 1 stop bit and no parity.

• Select "PRINTER" to enable the printer port.

The built-in printer is connected to the PRINTER port.

To disable the built-in printer, use the

INTERNAL/EXTERNAL switch on the back of the 2000-

610-02 Indicator.

• Select "PC" to enable the OPTIONAL COM port.

Selecting Printer/PC Ports

1. Press CAL to enter calibration mode.

— LED display reads the following with the least

significant digit (LSD) flashing:

CODE0000

2. Use left and right DIGIT to select the appropriate digits

and up and down VALUE to increment or decrement

each digit between "0" and hexidecimal "F," program the

number "2000" into the LED display.

Advertising