5-4

Calibration

2. Remove separated adaptor from calibration block.

3. Attach transducer to the calibration block so that red

mark on transducer lines up with white dot in left bottom

corner on calibration block. Tighten six 1/2"-13 hex head

bolts to 75 ft lb torque.

4. Attach the 1-1/4" drive adaptor plate using six 1/2"x13

hex head bolts, each torqued to 75 ft lb using the torque

wrench, and 3/4” open end adaptor, supplied.

5. After calibration, re-install the internal drive plate and

bottom plate using appropriate 1/2"x13 screws, torquing

them to 75 ft lb. Also re-install the separated adaptor to

the calibration block.

Appropriate hangers and weights are listed in the calibration

check point table in this chapter. Notice that the weight of the

hanger is included in all measurements. The hanger must be

removed to set ZERO TARE on the Indicator.

4-in-1 Transducer Mounting Detail for Calibration

The 4-in-1 transducer mounts upright with integral transducers

in the vertical position. This transducer may be mounted to

a:

• Wall using four 1/4" bolts torqued to 10 ft lb, or

• Sturdy table using a right angle bracket. Refer to Figure 5-4.

Figure 5-4: 4-in-1 Transducer Mounting Detail

The 4-in-1 transducer may also be mounted to the calibration

stand, shown in Figure 5-2, using the fixture multiple

transducers P/N 2500-500-02. For this mounting configuration

the two slide knobs must be loosened and four 5/16-18 bolts

attached to secure the 4-in-1 transducer to the bracket

assembly.

NOTE: Transducers the with four bolt flange, tighten the

four 3/8”-16 hex bolt to 45 ft lb.

Amendment for User’s Manual: December 08, 2006.