CDI Torque MULTITEST Government User Manual

Page 79

Advertising
background image

6-1

2000-600-02

Manual Loader

The 2000-600-02 Manual Loader is designed as an integral

component of the CDI MULTITEST Torque Calibration System

providing a precision output torque of up to 2000 ft lb while

requiring only 8 ft lb input.

Application

The loader is a laboratory grade instrument used for testing and

calibrating all torque wrenches, drivers, torque multipliers, non-

impact pneumatic and electric nut runners. This equipment also

can be mounted on a fixture for testing and calibrating cable

tensiometers and compression or tension gauges.

Power

To activate the limit solenoids that protect the transducers from

overload, the loader requires AC power, 115/230VAC, 50 to 60

Hz, 3.15 amps.

Do not use this equipment with the power off.

Always turn on the indicator and loader so the

torque and load values are indicated on the display.

The safety relays only work when the power is on.

Transducer Rotation

The manual hand crank provides up to 20° of transducer

rotation in both clockwise (CW) and counter clockwise (CCW)

directions.

MULTITEST Torque Calibration System

Components included in the 2000-600-02 System are:

• 2000-610-02 Indicator,

• 2000-600-02 Loader, and

• 2000-100-02 Roll Cabinet.

When the 2000-600-02 Manual Loader interfaces with the 2000-

610-02 Torque/Force Indicator and the MULTITEST series of

torque transducers, the MULTITEST Torque Calibration system

provides precision loading, monitoring and data acquisition of

torque, force and angle parameters within +/- 0.25% accuracy.

System capabilities include, data storage, retrieval, statistical

analysis and automatic downloading to a printer or computer.

6

WARNING

Advertising