Angle only – CDI Torque MULTITEST Government User Manual

Page 63

Advertising
background image

4-15

Using the MULTITEST Toque Calbration System

Angle Only

The MULTITEST accepts input from common bi-directional

incremental rotary encoders. The Indicator interfaces to any

encoder resolution (counts per revolution) and can be

programmed to display angle in degrees of rotation.

For additional information on setting up angle encoders and

measurement modes, refer to Chapter 3—Set Up and

Programming in this manual.

1. Press MODE to select ANGLE.

— LED displays torque.

— LCD displays degrees of rotation at the lower left.

Figure 4-6: Angle Mode Display

2. Install the angle encoder between the fastener and an

appropriate wrench or breaker bar. Fix the encoder

reference arm, strap, clamp or magnet to the work.

3. Press ANGLE RESET to zero the angle display.

4. Rotate the fastener while monitoring the angle on the

LCD display. The SET HIGH limit momentarily sounds

an audible tone at angle preset coincidence (when the

measured angle matches the SET LOW angle).

5. Press ANGLE RESET to clear the angle display.