5-9

Calibration

5. Enter the following code into the display using the

DIGIT right and left keys and the up and down VALUE

keys:

CODE5222

6. Press ENTER.

— The LED display reads:

SELECT

— The LCD display reads:

Calibrate CW

7. Press ENTER.

— The LED display reads a flashing number in the

LSD.

— The LCD display read:

Enter no. of transducers

8. Using the up and down VALUE keys, enter “1” for a

single transducer or “4” for the 4-in-1 transducer.

— If you enter "1" skip to step 11.

9. Press ENTER. The LED display reads a flashing number

in the LSD and the LCD display reads:

Select transducer

10. Using the up and down VALUE keys, enter the number of

the 4-in-1 transducer to calibrate. Possible values are:

1

=

50 in oz

2

=

400 in lb

3

=

1000 in lb

4

=

250 ft lb

11. Press ENTER.

— The LED display banners:

ENTER FS

followed by the last full scale calibration point for the

particular transducer installed or selected with its

MSD flashing.

— The LCD display reads:

ENTER FS transducer

[units]

12. Using the up and down VALUE keys and the left and

right DIGIT keys, enter the full scale torque capacity of

the transducer in the engineering units selected.

13. Press ENTER.

— The LED display banners:

ZERO TARE

Four zeros follow shifting alternately across the

display.

— The LCD display reads:

CALIBRATION

<ZEROTARE> CONT.