4-13

Using the MULTITEST Toque Calbration System

How to Use Analog Output (Rear Panel BNC

Connector)

Analog Output on the MULTITEST Indicator provides a real time

voltage level that is directly proportional to the torque/force

measurement applied to the transducer. It is useful for direct

driving equipment such as analog plotters and chart recorders,

or interfacing to a computer/controller with analog to digital

(A/D) data acquisition capability. Maximum output loading must

be less than 5 mA.

The output varies between approximately +1.8 volts, full

clockwise (CW) torque/force applied and approximately -1.8

volts, full counter clockwise (CCW) torque/force applied. Zero

offset is within ±100 mv. Linearity is better than ±1% of

reading.

When used with the 4-in-1 transducer, zero output

varies with the scan of the four transducers. Select

one of the four by applying at least 5% of its full range

torque to it. Press ENTER to return to scanning.

1. To program the MULTITEST Indicator, refer to Chapter

3—Set Up and Programming. Use SET HIGH and SET

LOW limit alert, data STORE, display CLEAR, printer

SEND and STATistical analysis features.

2. Press MODE to select the TRACK mode.

3. Connect the recording or monitoring equipment to the

ANALOG OUTPUT port using a BNC coaxial connector.

4. With no torque applied, measure the ANALOG OUTPUT

voltage and adjust or program your equipment for zero.

5. Apply full range torque to the transducer in the CW

direction as shown on the MULTITEST Indicator LED

display.

6. Measure the ANALOG OUTPUT voltage and adjust or

program your equipment to display the full range torque

value.

7. Apply full range torque to the transducer in the CCW

direction as shown on the MULTITEST Indicator LED

display.

8. Measure the ANALOG OUTPUT voltage and adjust or

program your equipment to display the full range torque

value.