5-2

Calibration

Mounting Details

Single Transducer Mounting Detail for Calibration

Single transducer calibration requires the use of a reaction

stand and transducer calibration block.

As shown in Figure 5-2, the transducer is secured to the

calibration block with two quick release pins.

• Both pins must be installed for safety and stability of the

mount.

• The white mark on the transducer is aligned in the up

position.

Figure 5-2: Single Low Torque Transducer Mounting Detail

Torque Under 250 ft lb

Mount the calibration block at the center of the stand with two

1/2-13 socket head cap screws with test bars that are 10" or

shorter. Install these screws using the supplied torque wrench

and 3/8-3/8 socket driver adapter. Use this configuration for up

to 250 ft lb of applied torque.

A 1/2" to 3/8" reducer is required when mounting the

10" arm to transducers having 3/8" drives.