3-14

Setup and Programming

B. Rotate the encoder shaft exactly one complete turn

(360 degrees). The upper right LCD display reads

the total number of counts for the encoder.

C. Push ENTER then CAL to return to the previous

mode.

To enter the resolution of a known encoder:

A. Press up VALUE to select "Encoder: Enter

Resolution" on the LCD display.

B. Press ENTER and, using up and down VALUE and

left and right DIGIT, enter the counts resolution of

the encoder.

C. Push ENTER then CAL to return the previous

mode.

4. Check the encoder count to degrees of rotation

conversion by pressing up VALUE to select "Encoder:

Check Resolution" on the LCD display.

5. Press ENTER and rotate the encoder shaft exactly one

complete turn.

— The LCD display reads:

Angle[0]= 0360.0

Figure 3-7: Encoder Count Display

6. Press ENTER.

7. Press CAL to return to measurements mode.

"Count[1]" and "Angle[1]" in the LCD display are future

options for use with a second encoder channel.