Setup and Programming

3-15

Setting Up Torque-Angle Measurement Modes

In using the torque-angle or "turn-of-the-nut" algorithm, a

fastener is run down to a specified clamping torque and then

rotated an additional number of angular degrees regardless of

the torque necessary to turn it. A typical fastener specification

might read: "15 ft lb & 50° rotation." For additional information,

refer to the Torque-Angle Tutorial Quick Reference card that

accompanies this manual.

When setting up torque-angle measurement modes:

• Angle measurements are not stored and cannot be printed.

• CLEAR, STORE and SEND must be set to MANUAL in the

ANGLE mode. For additional information, refer to

Programming AUTO CLEAR in this chapter.

• Torque SET HIGH must be set above the maximum torque

expected. For additional information, refer to Setting Up High

and Low Limits Alert in this chapter.

Setting Up the Torque/Angle Algorithm

When in the ANGLE mode, use SET LOW to preset the

MULTITEST Indicator to a clamping torque. At torque preset

coincidence (when the input torque matches the SET LOW

preset value), the MULTITEST Indicator momentarily sounds an

alert and automatically switches to measure angle from zero

degrees. Preset the maximum angle using the SET HIGH

function. A pulsating alert sounds at angle preset coincidence

(when the measured angle matches the SET HIGH maximum

angle preset value). SET HIGH torque preset also is active in

the angle mode to protect the wrench from over-torque.

Set Up Torque-Angle Algorithm

Use the following procedure to set up the torque-angle

algorithm.

1. Select ANGLE mode using MODE.

2. Press SET LOW.

— The LED display shows the previously programmed

clamping torque preset value.

— The LCD display reads:

Clamping torque - upper line

[units] - lower line

Figure 3-8: Clamping Torque Display