HI-6022 Field Probe

The ETS-Lindgren HI-6022 Field Probe

provides broadband frequency coverage

and wide dynamic range that satisfies the

demands of most test requirements.

The frequency response of the HI-6022 is

10 kHz to 1 GHz, and the dynamic range is

2 to 800 Volts per meter (V/m).

HI-6022

S

PECIFICATIONS

Dynamic Range:

2.0–800 Volts per meter (V/m)

Ranges:

Single

Typical Frequency Response:

10 kHz–1 GHz

• 10 kHz–30 kHz -2.5 / +0.5 dB

• 30 kHz–1 GHz ±1 dB

Linearity:

±0.5 dB

Isotropicity:

±0.5 dB

Overload Withstand:

> 1,500 V/m CW

Environmental

Operating Temperature:

10

°

C to 40

°

C

50

°

F to 104

°

F

Humidity:

5% to 95% relative humidity,

non-condensing

42 |

Battery-Operated Field Probes: HI-6022 Field Probe