HI-4422

O

PERATION

The HI-4422 is a battery-operated broadband radio frequency (RF) probe

designed for a variety of applications, including broadcast facilities, industrial

RF source and electromagnetic compliance (EMC) testing.

The HI-4422 measures field strength in each of three axes. It performs a vector

addition calculation on the readings and sends the result to the receiver through

a fiber optic cable. Data from each axis can be viewed individually, or can be

combined.

Limitations in system resolution may result in a non-zero reading when the

receiver is zeroed. If this occurs, it does not mean there is a problem with the

receiver, or that the readings are inaccurate. Probe linearity is specified as

+0.5 dB full scale. In addition, the variance of the A/D converter is +2 least

significant bits. When using the most sensitive range (10 V/m), these

specifications create the possibility that, under zero field conditions, the receiver

may display a non-zero value.

For a list and description of communication and information transfer protocols,

including command structure and probe commands, see Appendix E: Operating

Protocols on page 105.