Appendix d: series h-491198-36 battery charger, For nicd batteries – ETS-Lindgren HI-4422 Field Probe (Archived) User Manual

Page 99

Advertising
background image

Appendix D: Series H-491198-36 Battery Charger

for NiCd Batteries

The HI-4422 Field Probe contains a Nickel-Cadmium (NiCd) battery,

and uses the Series H-491198-36 Battery Charger.

S

AFETY

P

RECAUTIONS

Before operating the Series H-491198-36 Battery

Charger, see General Safety Considerations on

page ix.

I

NTRODUCTION

The H-491198-48 Nickel-Cadmium (NiCd)

Battery Charger is a dual power source

battery charger. It charges 3.6 volt

1400 mAH NiCd batteries and is powered

by 120-240 VAC line power or 12.5 VDC.

The H-491198-36 Battery Charger uses a

-(dV)/(dT) negative delta V technique to

determine when the battery is fully charged,

which is typically one hour. With this

technique, the charge state of the battery

has no effect other than shortening the

charge time.

Housed in a rugged enclosure, power enters the battery charger through a power

entry module, which contains the fuses, or an optional cigarette lighter plug

adapter. The front face of the battery charger displays LEDs that provide the

operating status. The battery charger connects to the device being charged

through a short cord terminated with a power jack.

An integrated circuit within the battery charger monitors the battery voltage and

controls the charging functions according to the charge state of the battery.

Series H-491198-36 Battery Charger for NiCd Batteries

|