S

YNCHRONIZING THE

B

ATTERY

C

HARGE

I

NDICATOR

The probe has internal circuitry to track the charge condition of the battery. NiMH

batteries have a sharp discharge-curve knee, which means there is little

indication that the battery is nearly discharged prior to a low battery condition.

The internal circuitry tracks the charging and discharging currents of the battery

and displays the calculated battery condition.

It is possible that the battery charge indication may lose synchronization with the

actual battery condition. A loss of synchronization may be indicated by a low

battery charge alert after the battery was recently fully charged. Loss of

synchronization may also be indicated by short battery life and premature low

battery alert.

To synchronize the battery charge indicator:

1.

Make sure the power switch on the probe is set to the OFF position or

the battery will not charge.

2.

Connect the battery charger to the electrical mains, and then connect

the charger output to the charger jack on the probe.

3.

Press the DISCHARGE button on the battery charger. The battery will

fully discharge, and then fully recharge. The total time required to

discharge and recharge is approximately seven hours.

4.

For best results, repeat the process once. However, it may be

necessary to repeat the process three times for proper

synchronization. The synchronization process may take up to

12 hours.

When completely charged, the battery indicator for the probe should indicate a

charge level of more than 90%.

D

ISCHARGING A

B

ATTERY

To discharge a battery:

1.

Make sure the power switch on the probe is set to the OFF position or

the battery will not charge.

2.

Connect the battery charger to the electrical mains, and then connect

the charger output to the charger jack on the probe.