320IS Installation Manual - Appendix A

53

7.6

Digital Filtering

The

320IS

uses RATTLETRAP

TM

digital filtering to reduce the effect of vibration on weight readings. Adjustable

threshold and sensitivity functions allow quick settling by suspending filter averaging, allowing the weight reading

to jump to the new value. Figure 7-2 shows the digital filter parameters on the CONFIG menu.

Figure 7-2. Digital Filtering Parameters on the Configuration (CONFIG) Menu

7.6.1

DFSENS and DFTHRH Parameters

The digital filter can be used to eliminate vibration effects, but heavy filtering also increases settling time. The

DFSENS (digital filter sensitivity) and DFTHRH (digital filter threshold) parameters can be used to temporarily

override filter averaging and improve settling time:

DFSENS specifies the number of consecutive scale readings that must fall outside the filter threshold

(DFTHRH) before digital filtering is suspended.

DFTHRH sets a threshold value, in display divisions. When a specified number of consecutive scale

readings (DFSENS) fall outside of this threshold, digital filtering is suspended. Set DFTHRH to NONE to

turn off the filter override.

7.6.2

Setting the Digital Filter Parameters

Fine-tuning the digital filter greatly improves indicator performance in heavy-vibration environments. Use the

following procedure to determine vibration effects on the scale and optimize the digital filtering configuration.

1. In setup mode, set the digital filter to 1. Set DFTHRH to NONE. Return indicator to normal mode.
2. Remove all weight from the scale, then watch the indicator display to determine the magnitude of vibration

effects on the scale. Record the weight below which all but a few readings fall. This value is used to

calculate the DFTHRH parameter value in Step 4.

For example, if a heavy-capacity scale produces vibration-related readings of up to 50 lb, with occasional

spikes to 75 lb, record 50 lb as the threshold weight value.

3. Place the indicator in setup mode and set the digital filter (DIGFLT) to eliminate the vibration effects on

the scale. (Leave DFTHRH set to NONE.) Reconfigure as necessary to find the lowest effective value for

the digital filter.

4. Calculate the DFTHRH parameter value by converting the weight value recorded in Step 2 to display

divisions:

threshold_weight_value / DSPDIV

In the example in Step 2, with a threshold weight value of 50 lb and a display division value of 5D:

50 / 5D

= 10.

DFTHRH should be set to 10DD for this example. See Figure 3-2 on page 24 for more information

on display divisions.

5. Finally, set the DFSENS parameter high enough to ignore transient peaks. Longer transients (typically

caused by lower vibration frequencies) cause more consecutive out-of-band readings, so DFSENS should

be set higher to counter low frequency transients.
Reconfigure as necessary to find the lowest effective value for the DFSENS parameter.

OFF

16

8

32

DIGFLT

128

64

256

8OUT

32OUT

16OUT

64OUT

DFSENS

DFTHRH

2OUT

128OUT

4OUT

2DD

NONE

5DD

20DD

10DD

50DD

200DD

100DD

250DD