EDP Commands

The IQ plus 210 indicator can be controlled by a

personal computer or remote keyboard connected to the

indicator serial port. Control is pro vided by a set of

EDP commands that can simulate front panel key press

functions, display and change setup parameters, and

perform reporting functions. The EDP port provides the

capability to print configuration data or to s ve that data

to an attached personal computer

. This section

describes the EDP command set and procedures for

saving and transferring data using the serial port.

5.1

The EDP Command Set

The EDP command set can be divided into fi e groups:

key press commands, reporting commands, the

RESETCONFIGURATION

special function

command, parameter setting commands, and transmit

weight data commands.
When the indicator processes an EDP command, it

responds with the message

OK

. The

OK

response

verifies that the command w as received and has been

executed. If the command is unrecognized or cannot be

executed, the indicator responds with

??

.

The following sections list the commands and

command syntax used for each of these groups.

5.1.1

Key Press Commands

Key press EDP commands (see T able 5-1) simulate

pressing the buttons on the front panel of the indicator.

5.1.2

Reporting Commands

Reporting commands (see T able 5-2) send specifi

information to the EDP port. These commands can be

used in both setup mode and normal mode.

5.1.3

The RESETCONFIGURATION Command

The RESETCONFIGURATION command can be used

to restore all configuration parameters to their de ault

values.
This command is equi valent to using the RESET

function on the DEFLT menu.

NOTE:

All load cell calibration settings are lost when the

RESETCONFIGURATION command is run.

5.1.4

Parameter Setting Commands

Parameter setting commands allo w you to display or

change the current v alue for a particular configuratio

parameter (Tables 5-3 through 5-8).
Current configuration parameter settings can b

displayed in either setup mode or normal mode using

the following syntax:

command<ENTER>

Most parameter values only can be changed in setup

mode. Use the follo wing command syntax when

changing parameter values:

command=value<ENTER>

where

value

is a number or a parameter v alue. Use no

spaces before or after the equal (=) sign. If you type an

incorrect command or v alue, the display reads

??

.

Changes to the parameters are saved as they are entered

but typically do not tak e effect until you e xit setup

mode.
For example, to set the motion band parameter to 5,

type the following:

MOTBAND=5D<ENTER>

Command

Function

KZERO

Press the ZERO button

KUNITS

Press the UNITS button

KPRIM

Display primary units

KPRINT

Send demand data to serial port

KRIGHTARROW

In setup mode, move right in the menu
(press the UNITS button)

KENTER

In setup mode, press the ENTER
(ZERO) button

KEXIT

In setup mode only, exit setup mode

Table 5-1. EDP Key Press Commands

Command

Function

DUMPALL

List all parameter values

VERSION

Write IQ plus 210 software version

P

Write current displayed weight. See
Section 6.2 on page 26 for more
information.

ZZ

Write current weight and annunciator
status. See Section 6.2 on page 26 for
more information.

RS

Reset the indicator

S

Write one frame of stream format

Table 5-2. EDP Reporting Commands