Setup and Service Information

3

2.0

Setup and Service Information

This section describes setup and service procedures for the IQ plus 2100 digital bench scale, including installation

and maintenance information, replacement parts lists, and assembly dra wings. See Sections 2.1 through 2.5 for IQ

plus 210 indicator information; see Sections 6.6 and 6.7 for bench scale information.

2.1

Enclosure Disassembly

The indicator enclosure must be opened to connect

cables after replacing the load cell, or when connecting

serial communications cables or digital inputs to the

scale.

The IQ plus 210 has no on/off switch.
Before opening the unit, ensure the
power cord is disconnected from the
power outlet.

Use a wrist strap to ground yourself and protect
components from electrostatic discharge (ESD)
when working inside the indicator enclosure.

This unit uses double pole/neutral fusing which
could create an electric shock hazard. Procedures
requiring work inside the indicator must be
performed by qualified service personnel only.

The supply cord serves as the power disconnect for
the IQ plus 2100. The power outlet supplying the
indicator must be installed near the unit and be
easily accessible

Ensure power to the indicator is disconnected, then

remove the scre ws that hold the backplate to the

enclosure body, then lift the backplate a way from the

enclosure and set it aside.

2.2

Cable Connections

The IQ plus 210 pro vides three cord grips for cabling

into the indicator: two for the power cord and load cell

cabling, the third for communications and digital input

cables. The free cord grip comes with a plug installed to

prevent moisture from entering the enclosure. If your

application requires serial communication or digital

input cabling, remo ve the plug and install cables as

described in Sections 2.2.4 and 2.2.5 on page 5.

2.2.1

Cable Grounding

Except for the power cord, all cables routed through the

cord grips should be grounded ag ainst the indicator

enclosure. Do the following to ground shielded cables:

• Use the lockw ashers, clamps, and k ep nuts

provided in the parts kit to install grounding

clamps on the studs adjacent to the cord grips.

Install grounding clamps only for cord grips that

will be used; do not tighten nuts.

• Route cables through cord grips and grounding

clamps to determine cable lengths required to

reach cable connectors. Mark cables to remo ve

insulation and shield as described below:

• For cables with foil shielding, strip insulation

and foil from the cable half an inch (15 mm)

past the grounding clamp (see Figure 2-1). Fold

the foil shield back on the cable where the cable

passes through the clamp. Ensure silv

er

(conductive) side of foil is turned outw ard for

contact with the grounding clamp.

• For cables with braided shielding, strip cable

insulation and braided shield from a point just

past the grounding clamp. Strip another half

inch (15 mm) of insulation only to expose the

braid where the cable passes through the clamp

(see Figure 2-1).

IMPORTANT!

For load cell cables, strip the

yellow shield wire 3/4" past the grounding

clamp. Fold wire back and secure between the

cable and clamp. Shield wire function is

provided by contact between the cable shield

and the grounding clamp.

• Route stripped cables through cord grips and

clamps. Ensure shields contact grounding clamps

as shown in Figure 2-1. Tighten grounding clamp

nuts.

• Finish installation using cable mounts and ties to

secure cables inside of indicator enclosure.

Figure 2-1. Grounding Clamp Attachment for Foil-Shielded

and Braided Cabling

Warning

Caution

Cord grip

Insulated cable

Foil (silver side out)

Grounding clamp

Shield wire (cut)

Length of foil before folding
back on cable insulation

Cut insulation here
for foil-shielded cables

Braid

Cut insulation here
for braided cables

NOTE: Install lockwashers
first, against enclosure,
under grounding clamp

Advertising