Installation

3

2.0

Installation

The IQ plus 2100 digital bench scale is designed for easy

setup and installation. All models are preconfigured an

weight calibrated before shipment, with the load cell

connected to the indicator. For simple, standalone scale

applications, hardware installation consists of attaching

the indicator, mounting assembly, and bench scale, then

leveling the scale.

NOTE:

The IQ plus 2100 is weight calibrated before

shipment from the factory, but recalibration at the
installation site is strongly recommended. Calibration by
a certified scale technician is required for Legal-for-Trade
applications.

2.1

Unpacking and Assembly

Immediately after unpacking, visually inspect the IQ

plus 2100 to ensure all components are included and

undamaged. The shipping carton should contain the

indicator, the bench scale, the indicator mount (tilt

stand, column, or attachment bracket), this manual, and

a parts kit. If an y parts were damaged in shipment,

notify Rice Lak e Weighing Systems and the shipper

immediately.
The parts kit contains the items listed below:

• Capacity label (PN 42350)
• One grounding clamp (PN 53075), e

xternal

tooth lock washer, (PN 15133), and kep nut (PN

14626) for digital input or serial cable shield

grounding against the enclosure.

• Two wing knobs with n ylon washers for

attaching the indicator to the mounting

assembly, four rubber tilt stand feet (for tilt

stand option) or bench scale feet, jam nuts, lock

washers, and cap scre ws (for column and

attachment bracket options). See Section 6.7 on

page 28 for replacement part numbers.

To install, attach the mounting assembly to scale base,

then assemble the indicator to the mounting assembly .

Place the bench scale in the desired location then lift off

the scale platter and locate the b ubble level. Adjust the

corner feet until the scale is le vel and all feet are in

contact with the support surf ace. Tighten the jam nuts

against the scale to lock-in the level adjustment.
Plug power cord into power outlet to begin weighing.

The supply cord serves as the power
disconnect for the IQ plus 2100. The
power outlet supplying the indicator
must be installed near the unit and be
easily accessible.

2.2

Factory Setup

The IQ plus 2100 is preconfigured and weigh

calibrated at the factory before shipment. Table 2-1 lists

the default graduations and count-by configuration fo

each scale capacity.

If scale configuration must be changed, see Section 3.0

on page 4 for detailed configuration information. Se

Section 4.0 on page 14 for calibration instructions.

Caution

Scale

Capacity

Count by

Configuration Parameters

DECPNT

DSPDIV

5 lb

0.001 lb

888.888

1D

10 lb

0.002 lb

888.888

2D

25 lb

0.005 lb

888.888

5D

50 lb

0.01 lb

8888.88

1D

100 lb

0.02 lb

8888.88

2D

2.5 Kg

0.0005 Kg

88.8888

5D

5 Kg

0.001 Kg

888.888

1D

10 Kg

0.002 Kg

888.888

2D

25 Kg

0.005 Kg

888.888

5D

50 Kg

0.01 Kg

8888.88

1D

NOTE: All capacities are preconfigured for 5000 graduations
(GRADS parameter on CONFIG menu set to 5000).

Table 2-1. Default Scale Configurations