Introduction

1

About This Manual

This manual is intended for use by service technicians

responsible for installing and servicing IQ plus

®

2100

digital bench scales.
Configuration and calibration of the indicator can be

accomplished using the indicator front panel k eys, the

EDP command set, or Version 2.3 or later of the

Revolution

configuration utility. See Section 3.1 on

page 10 for information about configuration methods

Some procedures described in this
manual require work inside the indicator
enclosure. These procedures are to be
performed by qualified service personnel
only

.

Authorized distributors and their emplo yees

can view or download this manual from the

Rice Lake Weighing Systems distributor site

at

www.rlws.com

.

1.0

Introduction

The IQ plus 2100 digital bench scale consists of an IQ

plus 210 digital weight indicator and an RL2100 bench

scale. The indicator can be mounted to the bench scale

using the 12-inch column or the attachment brack et, or

can be separately mounted on the tilt stand.
The IQ plus 210 is a single-channel digital weight

indicator housed in a NEMA 4X/IP66-rated stainless

steel enclosure. The indicator front panel consists of a

large (.8 in, 20 mm), six-digit, se ven-segment LED

display, seven LED annunciators, and tw

o piezo

switches used to zero the scale and select the displayed

units. Features include:

• Drives up to four 350

load cells

• Supports 4- and 6-wire load cell connections
• Two configurable digital input
• Single serial port supports full duple

x, RS-232

communications at up to 9600 bps

• Available in 115 VAC and 230 VAC versions

The RL2100 bench scale is a vailable in the follo wing

configurations

• 5, 10, and 25 lb (2.5, 5, and 10 Kg) capacities with a

10" x 10" (25cm x 25cm) platform

• 50 and 100 lb (25 and 50 Kg) capacities with a 12" x

12" (30cm x 30cm) platform

The RL1042 anodized aluminum load cell is standard

in the RL2100 bench scale; the RL1380 stainless steel

load cell is a vailable as an option. A stainless steel

clamshell enclosure is also a vailable for load cell

protection.

1.1

Operating Modes

The IQ plus 210 has two modes of operation:

Normal (weighing) mode

Normal mode is the “production” mode of the

indicator. The indicator displays the gross weight,

and the LED annunciators described on page 2

indicate scale status and the type of weight v alue

displayed. Once configuration is complete and

legal seal is affi ed to the back of the indicator, this

is the only mode in which the IQ plus 210 can

operate.

Setup mode

Most of the procedures described in this manual

require the indicator to be in setup mode, including

configuration and calibration

To enter setup mode, remove the large fillister hea

screw from the enclosure backplate. Insert a

screwdriver or a similar tool into the access hole

and press the setup switch once. The indicator

display changes to show the word

CONFIG

.